Software error. Corrupt flash memory.
Meaning: A software error occurred.
Stored data in the flash memory is corrupt.
Reason 1: A bug in the software.
Cure 1: Restart digitizer to return IP into cassette and release cassette if
possible.
Remove jammed IP and cassette manually if restart not successful.
Install the newest digitizer software with option format.
The processing station is not available.
Meaning: The image processing station is not available.
The processing station is not ready to receive the image.
Reason 1: The processing station software is not running properly.
Cure 1: Check processing station software. Restart processing station.
Reason 2: The Ethernet connection is not working properly.
Cure 2: Check Ethernet connection.
Reason 3: Wrong digitizer or destination setting.
Cure 3: Connect to service interface via browser and check the digitizer and
destination settings (network addresses and names, AE titles, ports).
The image destination is not known.
Meaning: The Digitizer cannot send the data to the image processing
application.
The configured destination is unknown.
Reason 1: Wrong Digitizer or destination setting.
Cure 1: Connect to service interface via browser and check the Digitizer and
destination settings (network addresses and names, AE titles, ports).
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 43 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents