AGFA CR Series Service Manual page 180

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
CODE
ARC2602
ARC2603
ARC2700
ARC2800
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Erasure Unit reports LED at standby current error.
Meaning: The 'Erasure Unit' could not be switched off successfully.
LED currents are not OK. The LEDs may still be on!
Reason 1: The 'Power Board' is defective driver or LED sensoric fails.
Cure 1: Replace 'Power Board'.
Erase failed.
Meaning: The Erasure procedure failed in active operation.
Reason 1: 'Erasure Unit' or 'Cable Power Board to Erasure Unit' defective.
Cure 1: Restart digitizer to get detailed selftest result.
Incorrect Power Board version.
Meaning: Wrong hardware version for 'Power Board' detected during start-up.
This type of digitizer needs a 'Power Board' with increased power output for the
erasure unit.
Reason 1: 'Power Board' has been mounted, which is incompatible with this
type of digitizer.
Cure 1: Replace 'Power Board' with compatible version as described in the
spare parts list.
System time error detected.
Meaning: The realtime clock is not usable.
The coin cell battery for the realtime clock at the 'PMI Board' is empty.
The system time is needed to generate unique identifiers as required by
DICOM.
Reason 1: The battery is disconnected or got loose.
Cure 1: Check proper mounting of the coin cell battery at place B2 on the 'PMI
Board'.
Check time setting via service tool at item Digitizer-Settings in the webbrowser.
Restart digitizer.
Reason 2: Battery is flat. Battery voltage is too low.
Cure 2: Replace 'PMI Board'.
Check time setting via service tool at item Digitizer-Settings in the webbrowser.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 32 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents