Edition 1, Revision 2
03-2013
SW error at the image processing station.
Meaning: Software error in the processing station software.
The IP cannot be scanned because the IP status is wrong.
Digitizer rejects the scan because the processing station did not set the IP
status to 'exposed'.
Normally the IP status is set to 'exposed' by pressing the 'ID' button on the
processing station.
Reason 1: Problem with Digitizer configuration.
Cure 1: Check if digitizer is configured correctly on the processing station.
Reason 2: Problem with software installation at the image processing station.
Cure 2: Check processing station software installation.
Reinstall the image processing station software.
Use the latest software version.
The image destination is invalid.
Meaning: The Digitizer cannot send the data to the image processing
application.
The configured destination unknown by the Digitizer or set to invalid values.
Reason 1: Wrong Digitizer or destination setting.
Cure 1: Connect to service interface via browser and check the Digitizer and
destination settings (network addresses and names, AE titles, ports).
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 44 of 46