Software error.
Meaning: A software error occurred. Digitizer cannot proceed the current
action.
Reason 1: A bug in the software at the current action.
Cure 1: Restart digitizer to return IP into cassette and release cassette.
Install the newest digitizer software.
Reason 2: A bug in the software causing system stop. Restart not successfully.
Cure 2: Clear IP jam manually. Install the newest digitizer software.
Missing software functionality.
Meaning: The installed software does not support this asked functionality.
Reason 1: Software version is too old.
Cure 1: Install the newest digitizer software.
Software error
Meaning: A software error is causing system stop.
Reason 1: A bug in the software causing system stop. Restart not successfully
Cure 1: Clear IP jam manually. Install the newest digitizer software.