DD+DIS155.11E
CODE
ARC1300
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Decoded Barcode data is invalid.
Meaning: Start marks of the Barcode Label successfully found, but the
decoded data content is invalid.
Reason 1: Barcode Label is not clean.
Cure 1: Check for a clean Barcode Label on the IP.
Reason 2: Ambient Light is too bright.
Cure 2: Protect the digitizer from direct sunlight.
Reason 3: Cleaning Brush is not in park position.
Cure 3: Bring the Cleaning Brush outside into the proper park position.
Reason 4: Barcode Label on the IP defective.
Cure 4: Try it again with another IP.
Replace the defective 'Image Plate MD 1.0'.
Reason 5: 'PMT with Light Collector' defective.
Cure 5: Replace 'PMT with Light Collector'.
Reason 6: 'Optic Module' defective.
Cure 6: Replace 'Optic Module'.