(38) If the installation wizard pops up:
Select <cancel>
(39) Select service menu: <Error
Viewer Installation>.
(40) Click:
<ErrorViewerInstaller.msi>
(41) Click <run> to start the
installation.
(42) Follow the further instructions to
install the Error Viewer.
(43) Copy the application license file
(ALF) to the NX workstation
(e.g. directory
C:\AGFA\Healthcare\NX\
Configuration\Licensing).
(44) Double-click the Licensing
Manager icon on the desktop.
(45) Select Add License and browse
to the folder where the ALF file is
stored.
(46) Wait a few seconds until the
message appears: The licenses
have been updated.
(47) Close the License Manager.