AGFA CR Series Service Manual page 157

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
A problem occurred when trying to display an ErrorMessage on the image processing station. The digitizer
cannot communicate anymore with the ErrorViewer.
1. Check if the ErrorViewer is running properly.
2. Check the network connection (cabling).
3. Check the digitizer and destination settings (IP addresses, ports).
Tuesday, 04-Sep-2012 15:39:09
Item
Type Number:
Subtype Number:
Serial Number:
Manufacture Date:
Installation Date:
Software Version:
Errorcode:
Name:
Short Description:
Description:
ID of User Message:
User Message:
IP Address:
Subnet Mask:
Gateway:
Hostname:
AE Title:
IP Address of Destination: 192.192.192.193
Port:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Value
01-Jan-1970
01-Jan-1970
unknown
4313
OPTICS_BOARD_CONNECTOR
Optics board not detected
Meaning: Optic board not detected. Reason: Cable broken/not connected or
optic board defective. Cure: 1) Check optic module cabling. 2) Replace optic
module cabling, if error persists. 3) Replace optic module, if error persists.
3
Ein Fehler ist aufgetreten. Bitte zur Problembehebung an Kundendienst
wenden.
192.192.192.100
255.255.255.0
192.192.192.1
DefaultDig
CR10MUC
13001
Figure 2: Example for file ErrorInformation.html
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Chapter 3.3 / 9 of 46
Agfa Company Confidential