AGFA CR Series Service Manual page 276

Digitizer
Table of Contents

Advertisement

DD+DIS170.11E
The device is not foreseen for installation in the patient vicinity. Therefore the
equipment must be placed outside the patient vicinity according to the locally valid
regulation (UL 60601-1 or EN 60601-1).
The patient vicinity is defined as follows:
UL 60601-1:
o 1.83 m (6 feet) beyond the perimeter of the bed (examination table, dental chair,
treatment booth, and the like)
o Vertical 2.29 m (7.5 feet) above the floor
EN 60601-1:
o 1.5 m (4.9 feet) beyond the perimeter of the bed (examination table, dental
chair, treatment booth, and the like)
o Vertical 2.5 m (8.2 feet) above the floor
Other equipment that is in contact with the patient is not connected directly
to the Agfa system without additional protective measures.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
Patient vicinity
R = 1.5 m / 4.9 feet (EN 60601-1) or
h = 2.5 m / 8.2 feet (EN 60601-1) or 2.29 m / 7.5 feet (UL 60601-1)
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 11 / Page 22 of 24
Type 5151 / 100 / 110 / 200 / 210
R
h
1.83 m / 6 feet (UL 60601-1)
Figure 12
Installation Planning
Patient vicinity
patient_vicinity.cdr
Agfa Company Confidential
1
2
3