AGFA CR Series Service Manual page 177

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
CODE
ARC2410
ARC2411
ARC2412
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
IP transport fails reference sensor position.
Meaning: Drawer does not reach the 'IP Drawer Light Sensor' position,
while IP is transported by 'IP Drawer Module'.
Reason 1: Image plate damaged.
Cure 1: Check Image plate. Replace IP.
Reason 2: Image plate blocked by something.
Cure 2: Remove blocked IP. Check if IP transport path is free.
Reason 3: 'IP Drawer Module' not working correctly.
Cure 3: Restart digitizer to get selftest results. Check 'IP Drawer Module' with
cables.
Replace defective part 'Y-Cable Power Board to Drawer/Cass. LS',
'IP Drawer Light Sensor', 'IP Drawer Module'.
Reason 4: 'Power Board' defective.
Cure 4: Check 'Power Board'. Replace 'Power Board'.
Selftest error of 'IP Drawer Light Sensor'.
Meaning: Selftest detects 'IP Drawer Light Sensor' of 'IP Drawer Module'
disconnected.
Reason 1: Reference sensor is disconnected.
Cure 1: Check cable and connectors to 'IP Drawer Light Sensor'.
Reason 2: Sensor defective.
Replace 2: Replace 'IP Drawer Light Sensor'.
Reason 3: 'Power Board' defective.
Cure 3: Replace 'Power Board'.
Selftest error 'IP Detection Light Sensor'.
Meaning: Selftest detects 'IP Detection Light Sensor' disconnected.
Reason 1: 'IP Detection Light Sensor' is disconnected.
Cure 1: Check cable and connectors to 'IP Detection Light Sensor'.
Reason 2: 'IP Detection Light Sensor' defective.
Cure 2: Replace 'IP Detection Light Sensor'.
Reason 3: 'Power Board' defective.
Cure 3: Replace 'Power Board'.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 29 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents