AGFA CR Series Service Manual page 176

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
CODE
ARC2408
ARC2409
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
IP Transport error at position of 'IP Detection Light Sensor'.
Meaning: 'IP Detection Light Sensor' has wrong status for object inside.
Expected status is sensor-free.
Reason 1: 'IP Detection Light Sensor' is covered by something.
Cure 1: Check if 'IP Detection Light Sensor' is mounted correctly and not
covered by something.
Reason 2: The cable is defective or disconnected.
Cure 2: Check cables and connectors in the wiring to 'IP Detection Light
Sensor'.
Replace 'Cable Power Board to IP Detection LS'.
Reason 3 'IP Detection Light Sensor' defective.
Cure 3: Replace 'IP Detection Light Sensor'.
Reason 4: 'Power Board' defective.
Cure 4: Check 'Power Board'. Replace if defective.
Motor movement error of 'IP Drawer Module'.
Meaning: Any standard movement of 'IP Drawer Module' not finished
successfully.
Reason 1: Software problem.
Cure 1: Load new Software.
Reason 2: 'Power Board' defective.
Cure 2: Check 'Power Board' and replace if defective.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 28 of 46
Agfa Company Confidential