AGFA CR Series Service Manual page 181

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
CODE
ARC2801
ARC2802
ARC2804
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Connection between 'PMI Board' and 'PMT with Light Collector' failed.
Meaning: At the 'PMT with Light Collector' module the tube electronic board
detection failed.
Reason 1: Cable defective or disconnected.
Cure 1: Check cables and connectors between 'PMT with Light Collector' and
'PMI Board'.
Replace 'Data Cable PMI Board to Tube Board'. Replace 'Signal Cable PMI
Board to Tube Board'.
Reason 2: 'PMT with Light Collector' is defective.
Cure 2: Replace 'PMT with Light Collector'.
Reason 3: 'PMI Board' defective.
Cure 3: Replace 'PMI Board'.
Connection between 'PMI Board' and 'Optic Module' failed.
Meaning: 'Optic Module' electronic board not detected.
Reason 1: Cable defective or disconnected.
Cure 1: Check cables and connectors between 'PMI Board' and 'Optic Module'.
Replace 'Cable PMI Board to Optic Module'.
Reason 2: 'Optic Module' is defective.
Cure 2: Replace 'Optic Module'.
Reason 3: 'PMI Board' defective.
Cure 3: Replace 'PMI Board'.
Flash memory error at 'PMI Board'.
Meaning: The internal flash memory of the 'PMI Board' is corrupted or full.
Reason 1: Due to a software error the flash memory was corrupted or is full.
Cure 1: Install the newest software version with formatting of the flash memory.
Restore previously saved backup data.
Reason 2: Flash memory on 'PMI Board' is defective.
Cure 2: Replace 'PMI Board'.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 33 of 46
Agfa Company Confidential