AGFA CR Series Service Manual page 175

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
CODE
ARC2404
ARC2407
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Motor error at the 'IP Drawer Module'.
Meaning: During motor selftest of 'IP Drawer Module' an error occurred.
Reason 1: The cables are defective or disconnected.
Cure 1: Check cables and connectors between 'Power Board' and motor of 'IP
Drawer Module'.
Replace 'Y-Cable Power Board to Motors'.
Reason 2: The motor of 'IP Drawer Module' is defective.
Cure 2: Replace the 'IP Drawer Module'
Reason 3: 'Power Board' defective.
Cure 3: Replace 'Power Board'.
Reference at the 'IP Drawer Module' cannot be detected.
Meaning: The motor drive of the 'IP Drawer Module' does not reach
reference position controlled by 'IP Drawer Light Sensor'.
Reason 1: 'IP Drawer Light Sensor' is out of position.
Cure 1: Bring 'IP Drawer Light Sensor' in correct position.
Reason 2: The cable is defective or disconnected.
Cure 2: Check cables and connectors to 'IP Drawer Light Sensor' of 'IP Drawer
Module'.
Replace 'Y-Cable Power Board to Drawer/Cass. LS'.
Reason 3: Sensor defective.
Cure 3: Replace 'IP Drawer Light Sensor'.
Reason 4: 'IP Drawer Module' motor drive does not move at all or with difficulty.
Cure 4: Check motor drive of 'IP Drawer Module' for proper work and replace if
damaged.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 27 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents