DD+DIS155.11E
(8)
A screen with instructions is
displayed.
Enter the exposed cassette in the
digitizer: Scanning starts.
(9)
Wait until scanning is finished.
A static red status LED indicates
scanning.
A static green status LED indicates
that scanning is finished: The
shading calibration curve is
displayed.
IMPORTANT:
If a message pops up which indicates a failed shading calibration (e.g. "Underexposed
- too low pixel values ...) refer to section 2.5 for possible remedies.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.6 / Page 8 of 10