AGFA CR Series Service Manual page 189

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
CODE
ARC4007
ARC4008
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Scanning unit operating conditions error.
Meaning: The scanning process failed.
The Slow Scan Motor has already stopped.
ICC (Image Capture Control) is still waiting for end/begin of the scan, i.e. the
required number of pixels has not been stored yet.
Reason 1: The cables to 'PMT with Light Collector' are defective or
disconnected.
Cure 1: Restart Digitizer to get relevant selftest result.
Try to scan an IP again to test the PMT.
Check cables and connectors between 'PMI Board' and 'PMT with Light
Collector'.
Replace 'Data Cable PMI Board to Tube Board'.
Replace 'Signal Cable PMI Board to Tube Board'.
Reason 2: The cable to 'Optic Module' is defective or disconnected.
Cure 2: Check cable and connectors between 'PMI Board' and 'Optic Module'.
Replace 'Cable PMI Board to Optic Module'.
Reason 3: The 'PMT with Light Collector' is defective.
Cure 3: Replace 'PMT with Light Collector'.
Reason 4: 'PMI Board' defective.
Cure 4: Replace 'PMI Board'.
Reason 5: The 'Optic Module' is defective.
Cure 5: Replace 'Optic Module'.
Error information and storage problem while ErrorViewer is disabled.
Meaning: Could not save error information file and ErrorViewer is not
accessible.
Reason 1: USB flash memory not inserted or has no free memory left.
Cure 1: Insert a USB flash memory with 10 kBytes free memory left.
Reason 2: USB flash memory is write protected or does not allow write access.
Cure 2: Remove write protection or use an USB flash memory with 10 kBytes
free memory left.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 41 of 46
Agfa Company Confidential