IP-transport error.
Meaning: Movement of the Handling has not finished successfully.
The control reached an unexpected situation.
Check the description of the Error Code for more specific information.
Reason 1: Error during operation.
Cure 1: Restart digitizer to perform the startup selftest procedure.
Watch out for messages, e.g. if an IP has to be removed.
Reason 2: Software problem.
Cure 2: Load new Software.
Reason 3: 'Power Board' defective.
Cure 3: Check 'Power Board' and replace if defective.
No IP present at IP Sensor at the end of the tray.
Meaning: Transport of the IP to the 'IP Detection Light Sensor' not successful.
At the IP Detection Light Sensor no IP is detected after the scan movement.
Reason 1: Transport of the IP not successful.
Cure 1: Restart Digitizer.
Reason 2: IP has jammed.
Cure 2: Open Digitizer and remove jammed IP.
Unexpected IP found at the beginning of a new cycle.
IP present at IP Detection Light Sensor.
Meaning: At the last cycle the expected transport of the IP into the Cassette
was not successful.
At the 'IP Detection Light Sensor' an IP is detected at the start movement of the
new cycle.
Reason 1: Handling of the IP not successful.
Cure 1: Restart Digitizer.
Reason 2: IP has jammed.
Cure 2: Open Digitizer and remove jammed IP.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 18 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents