6 Instrument Functions

6.1 Setting the Channels Voltage and Current

NGL202 comes with two channels and NGL201 comes with single channel.
Toggle the respective channel key ([Ch1],[Ch2] ) on the front panel to select these
channels. When a channel is selected, the respective channel key illuminates.
Figure 6-1: Ch2 key illuminates when selected
Set output voltage and current
The R&S NGL200 adjust voltage values between 0 V to 20 V with a step size of 1 mV
and current values between 0 A to 6 A with a step size of 0.1 mA.
The setting of current value corresponds to the I
advisable to set the current limit prior to operating the instrument to prevent damage to
the load and instrument in the case of malfunction like short-circuit.
1. Press [Home] key.
The R&S NGL200 displays the home window.
2. Set voltage or current in the home window.
The R&S NGL200 displays an on-screen keypad to set value.
3. Enter the voltage or current value.
The R&S NGL200 displays the on-screen keypad to enter value.
4. Confirm value with the unit key.
5. Press the channel key ([Ch1] or [Ch2]) on the front panel.
The selected channel key is illuminated. See
6. Press the [Output] key on the front panel.
The R&S NGL200 output the set voltage of the selected channel and displays the
corresponding values in the home window.
For more information on the output mode, see
nels
User Manual 1178.8736.02 ─ 02.01
Output",
on page 40.
Instrument Functions
Setting the Channels Voltage and Current
of the respective channel. It is
max
Figure
6-1.
Chapter 6.2, "Activating the Chan-
39