Log:duration; Log:fname - R&S NGL200 series User Manual

Hide thumbs Also See for NGL200 series:
Table of Contents

Advertisement

®
R&S
NGL200
LOG:DURation <span>
LOG:DURation? [<span>]
Sets/Queries the duration of the data logging.
Setting parameters:
<span>
Query parameters:
<span>
Example:
LOG:FNAMe <filename>[,<loaction>]
LOG:FNAMe? [<location>]
Sets/Queries the filename and storage location for the logging function.
Setting parameters:
<filename>
<location>
Query parameters:
<location>
Example:
User Manual 1178.8736.02 ─ 02.01
numeric_values | MIN | MAX
numeric_values
Duration of the data logging captured in the range of 0 s to
3.49*10^5 s.
MIN
Minimum duration of the data logging captured at 0 s.
MAX
Maximum duration of the data logging captured at 3.49*10^5 s.
Default unit: s
MIN | MAX
Return the duration of the data logging.
LOG:DUR MAX
LOG:DUR? MAX -> 349000
Filename of the data logging function.
INT | EXT | DEF
INT
Internal memory.
EXT
USB stick.
DEF
Internal memory.
INT | EXT | DEF
Storage location for the logging function.
LOG:FNAM "Test01.CSV",INT
LOG:FNAM? -> "/INT/DATA/Test01.CSV"
Remote Control Commands
Data and File Management Commands
125

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ngl201Ngl202Ngl200Ngl-k102Ngl-k103Ngl-b105

Table of Contents