Figure A-1: Structure of the status QUEStionable register
CONDition
The CONDition register queries the actual state of the instrument. If you want to
query the constant voltage or current mode, you have to use the CONDition regis-
ter.
The CONDition register delivers a "1" (first bit set) in constant current mode (CC) and
a "2" (second bit set) in constant voltage mode (CV).
If the correct channel is selected and the red LED of the channel button lights up (CC
mode), the query of the CONDition register must deliver a "1".
Example:
STAT:QUES:ISUM1:COND?
EVENt
The EVENt status register is set (1) until it is queried. After reading (query), the
EVENt status register is set to zero.
User Manual 1178.8736.02 ─ 02.01
Additional Basics on Remote Control
Status Reporting System
139