[SOURce:]VOLTage:RAMP[:STATe] <state>
[SOURce:]VOLTage:RAMP[:STATe]?
Activates/Deactivates or Queries the Ramp function for the previous selected channel.
Parameters:
<state>
Example:
[SOURce:]VOLTage:RAMP:DURation <duration>
[SOURce:]VOLTage:RAMP:DURation?
Sets/Queries the duration of the voltage ramp.
Parameters:
<duration>
Example:
7.9 Data and File Management Commands
The DATA and HCOPy subsystem contains commands for managing the files in the
instrument and external USB stick.
User Manual 1178.8736.02 ─ 02.01
0 | 1
0
EasyRamp function will be deactivated.
1
EasyRamp function will be activated.
*RST:
0
INST OUT1
VOLT:RAMP ON
VOLT:RAMP? -> 1
EasyRamp function of Ch1 is activated
numeric | MIN | MAX | DEFault
numeric
Duration of the ramp function in seconds.
MIN
Minimum duration of the ramp function at 0.00 s.
MAX
Minimum duration of the ramp function at 10 s.
DEF
Default duration of the ramp function at 0.01 s.
Range:
0.01 to 10.00
*RST:
0.01
Default unit: s
VOLT:RAMP:DUR 4
VOLT:RAMP:DUR? -> 4
Duration of the ramp function is set at 4 s.