R&S
NGL200
Example:
Usage:
DATA:DELete <filename>[,<location>]
Deletes the logging file data values of the selected storage location and file name.
If no storage location is selected, the instrument uses the internal memory. Please
notice that the logging function has to be activated, if you want to use the manual trig-
ger mode (trigger via TRIG button). Without activating the logging function in manual
trigger mode, the instrument is not able to save a logging file internally or on the USB
stick.
Setting parameters:
<filename>
<location>
Example:
Usage:
DATA:LIST? [<location>]
Queries all saved logging files of the selected storage location.
User Manual 1178.8736.02 ─ 02.01
DATA:DATA? "LOG0001.CSV",EXT ->
0.019;2.0310;1.000;0.0000;1.000;0.0000;03:45:
29:376
0.020;2.0310;1.000;0.0000;1.000;0.0000;03:45:
30:375
0.020;2.0310;1.000;0.0000;1.000;0.0000;03:45:
31:374
0.020;2.0310;1.000;0.0000;1.000;0.0000;03:45:
32:373
0.019;2.0310;1.000;0.0000;1.000;0.0000;03:45:
33:372
External logging file (USB stick), count = 5.
Return as follows: U1[V];I1[A];U2[V];I2[A];U3[V];I3[A];Timestamp
Query only
Filename of the logging file data to be deleted.
INT | EXT | USB
INT
Internal memory
EXT
USB stick
DEF
Internal memory
DATA:DEL "LOG0001.CSV",EXT
Delete filename LOG0001.CSV from USB stick
Setting only
Remote Control Commands
Data and File Management Commands
121