Table 6-2: Trigger-out signals
Trigger-out parameters
Output channel
Digital Output Fault
Out 1
Out 2
Logging
Arb
DIO connector
The digital I/O connector is located below the GPIB connector, see
"Rear
The following shows the DIO connector and pins layout of the connector. See
ure 6-18
Figure 6-18: DIO connector (female socket front view)
User Manual 1178.8736.02 ─ 02.01
Trigger conditions
Output On
Output Off
Gated
CC, CV, CR, Protection and Sink
Trigger-in signal
Output On Ch 1
Trigger-in signal
Output On Ch 2
Trigger-in signal
Start triggered
Stop triggered
Panel",
on page 21.
and
Table
6-3.
Instrument Functions
Advanced Features
Descriptions
If a trigger is detected,
respective channel output of
the instrument turns on or
off.
If respective channel
output
modes,
protection event
or
sink mode is detected, a trig-
ger signal is sent out at pin
11 of the
DIO
connector.
If detected, a trigger signal is
sent out at pin 4 of the
DIO
connector.
If detected, a trigger signal is
sent out at pin 12 of the
DIO
connector.
If detected, the data logger
starts recording the instru-
ment.
If a trigger is detected,
respective channel starts as
a complete arbitrary signal or
steps through the arbitrary
signal for every trigger signal
detected.
The step time from the arbi-
trary signal is ignored in the
case when trigger condition
is set as "Stop triggered".
Chapter 4.2.1.2,
Fig-
56