// response: 60000
// ************************************************
// Link the electronic fuses of the channels logically
// ************************************************
INST OUT1
FUSE:LINK 2
// links the fus e of channel 1 with channel 2
FUSE:LINK?2
// queries the combined fuses of the selected channel
// ************************************************
// Unlink linked fuse
// ************************************************
FUSE:UNLink 2
FUSE:LINK?2
// queries the combined fuses of the selected channel
//response 0
FUSE:DELay:INITial
FUSE:DELay[:BLOWing]
FUSE:LINK
FUSE:TRIPped?
FUSE:UNLink
FUSE[:STATe]
FUSE:DELay:INITial <delay>
FUSE:DELay:INITial?
Sets/Queries an initial delay time for the fuse to take effect.
Parameters:
<delay>
Example:
User Manual 1178.8736.02 ─ 02.01
.......................................................................................................104
................................................................................................ 105
....................................................................................................................105
............................................................................................................ 105
................................................................................................................ 105
................................................................................................................106
numeric | MIN | MAX
numeric
Numeric value for linitial fuse delay.
MIN
Min value for lowe linitial fuse delay.
MAX
Max value for linitial fuse delay.
Range:
*RST:
Default unit: s
See
Example "Configuring fuses"
Remote Control Commands
0.00 to 60.00
0
on page 102.
Configuration Commands