7.4 Trigger Commands

The TRIGger subsystem contains the commands for signal triggering.
TRIGger:SLOPe
TRIGger:SLOPe <slope>
Sets/Queries the type of external trigger input.
Parameters:
<slope>
Example:

7.5 Configuration Commands

The following subsystems contain the commands for channel selection, voltage and
current settings for the instrument.

7.5.1 Channel Selection

The INSTrument:Select subsystem contains the commands for selecting the output
channels.
Each channel of the power supply is considered as separate "instrument", which is
required by the SCPI standard. Therefore, the SCPI commands use the INSTRument
node to select a channel.
You can only address the number of channels a device is equipped with, e.g. a maxi-
mum of two channels for the R&S NGL202, or one channel for the R&S NGL201.
User Manual 1178.8736.02 ─ 02.01
.............................................................................................................. 85
POSitive | NEGative
POSitive
Rising edge of the external trigger input.
NEGative
Falling edge of the external trigger input.
TRIGger:SLOPe POSitive
TRIGger:SLOPe? -> POS
Queries the type of EXTERNAL trigger input
Remote Control Commands