Figure 5-10: Current limit
CC mode
The current I
If I
out
constant and limited to I
decreases to almost zero with a short circuit. In CC mode, the font text in the channel
display area changes to red.
See
CR mode
To go into CR mode, the R&S NGL200 must operates in sink more where current flows
into the instrument and the "Constant Resistance" is enabled.
With "Constant Resistance" enable and configured, the R&S NGL200 can vary the
resistance in sink mode, this allows the R&S NGL200 to behave like an electronic load
which is useful in battery test application.
In CR mode, the font text in the channel display area changes to cyan. See
User Manual 1178.8736.02 ─ 02.01
Figure
5-2.
corresponds to the current setting adjustable in the instrument.
max
reaches I
, the instrument switches to CC mode, i.e. the output current remains
max
max
Figure
5-2.
even if the load increases. Instead, the output voltage V
Operating Basics
Output Modes
out
Figure
5-2.
38