5.3.3 Output and Channel Controls

Only applicable for R&S NGL202, these keys control the channel output settings of the
instrument.
Function keys
[Ch1], [Ch2]
[Output]

5.4 Power Derating

The NGL202 includes two identical channels with a continuous voltage range of 0 V to
20 V. The instrument provides a source of up to 6 A for voltage below 6 V and 3 A for
voltage range from 6 V to 20 V.
Similar to NGL202, the NGL201 provides a single channel wiith an output power of up
to 60 W.
Combination of the set voltage and current limit results in the following output perform-
ance graph.
Figure 5-9: Output performance graph

5.5 Output Modes

The R&S NGL200 includes three operating modes, i.e. the constant voltage (CV), con-
stant current (CC) and constant resistance (CR). The instrument switches automati-
cally between CV and CC depending on the connected load. In CR mode, the instru-
ment is not switched automatically into sink more. Instead, the instrument operates in
the CR mode
CV mode
Figure 5-10
remains constant while the current may increase to its maximum value I
connected load is increasing. In CV mode, the font text in the channel display area
changes to green.
User Manual 1178.8736.02 ─ 02.01
(if
configured) as soon as it switches from soure mode to sink mode.
shows that in the range of voltage regulation, the output voltage V
Description
Selects the respective channel for output.
Toggles the selected channel output on or off.
Operating Basics