Figure 6-16: Ramp dialog
4. Activate the "Enabled" menu item.
The R&S NGL200 enables the ramping function and displays the "Ramp" icon on
the selected channel status bar information.
5. Enter the required ramping time.
The R&S NGL200 displays an on-screen keypad to set the value.
6. Confirm value with the unit key.

6.4.3 Digital I/O

Digital I/O Pins Voltage Rating
Do not exceed the maximum voltage rating of the Digital I/O pins (5.25 V max) when
supplying voltages to the pins.
R&S NGL-K103 (3652.6385.02) option is required for the Digital I/O signals.
Option is not required for "User Button" and "*TRG" trigger-in signals.
With an optional digital I/O, you can apply a single trigger-in signal and control multi
trigger-out signals on the instrument. These trigger-out signals can be used to perform
function such as triggering the data logger to record instrument when a channel output
is active or protection event is used to trigger the digital out pin, which in turn can be
used for fuse linking between two NGLs.
See
User Manual 1178.8736.02 ─ 02.01
Figure 6-17
for an overview of the trigger IO system.
Instrument Functions
Advanced Features
54