10
ROUTe Subsystem

ROUTe:CHANnel:RANGe

Syntax
ROUTe:CHANnel:RANGe <value>, (@<ch_list>)
This command is used to set the voltage range of the specified analog
input channels.
ROUTe:CHANnel:RANGe? (@<ch_list>)
The query command returns a float value that represents the range in
volts of the specified analog inputs channels. Multiple responses are
separated by commas.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Item
Type
<value>
Float
<ch_list>
Numeric
88
Range of Values
U2331A: Unipolar mode: {10|5|4|2.5|2.0|1.0|0.5|0.4|0.1}
Bipolar mode: {10|5|2.5|2.0|1.25|1.0|0.5|0.25|0.2|0.05}
U2351A/U2352A/U2353A/U2354A/U2355A/ U2356A:
Unipolar/Bipolar mode: {10,5,2.5,1.25}
Single-ended mode:
U2351A/U2352A/U2353A/U2354A: 101-116
U2355A/U2356A/U2331A: 101-164
Differential mode:
U2351A/U2352A/U2353A/U2354A: 101-108
U2355A/U2356A/U2331A: 101-132
U2300A Series USB DAQ Programming Guide
Default Value
10
N/A