[SENSe:]COUNter:CLocK:POLarity

Syntax
[SENSe:]COUNter:CLocK:POLarity <mode>, (@<ch_list>)
This command is used to set the clock polarity settings of the general
purpose counter.
[SENSe:]COUNter:CLocK:POLarity? (@<ch_list>)
The query command returns a string value that represents the clock
polarity settings of specified counter channel of the general purpose
digital counter. Multiple responses are separated by comma.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Query Returned Format
<mode> ::= {AHI|ALO}
U2300A Series USB DAQ
Item
Type
<mode>
String
<ch_list>
Numeric
Range of Values
AHI: Active High
ALO: Active Low
301 through 302
SENSe Subsystem
11
Default Value
AHI
N/A
115