[Sense:]Voltage:stype - Agilent Technologies U2300A Series Programming Manual

Multifunction usb data acquisition
Hide thumbs Also See for U2300A Series:
Table of Contents

Advertisement

11
SENSe Subsystem

[SENSe:]VOLTage:STYPe

Syntax
[SENSe:]VOLTage:STYPe <mode>, (@<ch_list>)
This command is used to set the reference ground of the specified analog
input channel.
[SENSe:]VOLTage:STYPe? (@<ch_list>)
The query command returns a string value representing the reference
ground of the specified analog input channel. Multiple responses are
separated by comma.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Item
<mode>
<ch_list>
104
Type
Range of Values
String
DIFF: Differential mode
SING: Referenced single-ended mode (16/64 ch
common to ground system on board)
NRS: Non-referenced single-ended mode (16/64 ch
common to AISENSE pin)
Numeric
Single-ended mode:
U2351A/U2352A/U2353A/U2354A: 101-116
U2355A/U2356A/U2331A: 101-164
Differential mode:
U2351A/U2352A/U2353A/U2354A: 101-108
U2355A/U2356A/U2331A: 101-132
Default Value
SING
N/A