7
MEASure Subsystem
MEASure:DIGital
Syntax
MEASure:DIGital? (@<ch_list>)
The query command returns a numeric value representing the digital data
from the specified digital input port. Multiple responses are separated by
commas.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Remarks
This query requires a channel number to be stated.
Query Returned Format
Channel 501, 502: 0 to 255 (8- bits)
Channel 503, 504: 0 to 15 (4- bits)
Example
CONFigure:DIGital:DIRection INPut, (@501)
MEASure:DIGital? (@501)
See Also
CONFigure:DIGital:DIRection
60
?
Item
Type
<ch_list>
Numeric
Range of Values
Channel 501 through 504
U2300A Series USB DAQ Programming Guide
Default Value
N/A