CONFigure:DIGital:DIRection
Syntax
CONFigure:DIGital:DIRection <direction>, (@<ch_list>)
This command is used to configure the selected digital port and the
direction setting for input and output operations of the selected port.
CONFigure:DIGital:DIRection? (@<ch_list>)
The query command returns "INPut" or "OUTPut" for the specified
channels. Multiple responses are separated by comma.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Query Returned Format
<direction> ::= {INPut|OUTPut}
Example
CONFigure:DIGital:DIRection INPut, (@501)
CONFigure:DIGital:DIRection? (@501)
//Expected returned string is INPut
See Also
MEASure:DIGital?
MEASure:DIGital:BIT?
U2300A Series USB DAQ Programming Guide
Item
Type
<direction>
String
<ch_list>
Numeric
Range of Values
{INPut|OUTPut}
Channel 501 through 504
CONFigure Subsystem
Default Value
INPut
N/A
5
27