5
CONFigure Subsystem

CONFigure:TIMEbase:SOURce

Syntax
CONFigure:TIMEbase:SOURce <mode>
This command is used to set the timebase/clock source for both analog
inputs and outputs.

CONFigure:TIMEbase:SOURce?

The query command returns a string value, which represents the
timebase/clock source for both analog input and output channels.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Query Returned Format
<mode> ::= {INTernal|EXTernal|CCG}
Example
CONFigure:TIMEbase:SOURce EXTernal
CONFigure:TIMEbase:SOURce? //Expected returned string is EXTernal
28
Item
Type
<mode>
String
Range of Values
INTernal: Internal clock as the base time
EXTernal: External clock as the base time
CCG: Clock based on U2781A modular
instrument chassis
U2300A Series USB DAQ Programming Guide
Default Value
INTernal