[SENSe:]COUNter:ABORt

Syntax
[SENSe:]COUNter:ABORt (@<ch_list>)
This command aborts an initiated counter measurement in progress on the
specified counter channels. This command may be effective to abort a long
internally- gated or externally- gated measurement.
Supported Device
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Remarks
This command does not clear or invalidate any measurement in memory.
Example
[:SENSe:]COUNter:FUNCtion TOTalize (@301)
[:SENSe:]COUNter:TOTalize:INITiate (@301)
[:SENSe:]COUNter:ABORt (@301)
See Also
[SENSe:]COUNter:TOTalize:CLEar
108
Item
Type
<ch_list>
Numeric
Range of Values
301 through 302
Default Value
N/A
U2300A Series USB DAQ