[SENSe:]COUNter:TOTalize:UDOWn:DIRection

Syntax
[SENSe:]COUNter:TOTalize:UDOWn:DIRection <mode>,
(@<ch_list>)
This command is used to set the up or down counting process.
[SENSe:]COUNter:TOTalize:UDOWn:DIRection? (@<ch_list>)
The query command returns a string value that indicates the up or down
direction count direction of each specified counter channel. Multiple
responses are separated by comma.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Query Returned Format
<mode> ::= {UP|DOWN}
U2300A Series USB DAQ
Item
Type
<mode>
String
<ch_list>
Numeric
Range of Values
UP: Up counting
DOWN: Down counting
301 through 302
SENSe Subsystem
Default Value
UP
N/A