MEASure:COUNter:DATA?
Syntax
MEASure:COUNter:DATA? (@<ch_list>)
The query command returns a float value based on the SENSe:COUNter
command. This value is the counter value for specific counting mode on
specified counter channel. Multiple responses are separated by commas.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Remarks
This query requires a channel number to be stated.
Query Returned Format
<value>
Example
SENSe:COUNter:FUNCtion TOTalize, (@301)
SENSe:COUNter:FUNCtion FREQuency, (@301)
MEASure:COUNter:DATA? (@301)
See Also
[SENSe:]COUNter:FUNCtion
U2300A Series USB DAQ Programming Guide
Item
Type
<ch_list>
Numeric
Range of Values
301 through 302
MEASure Subsystem
Default Value
N/A
7