IEEE-488 Common Command
Remarks
• The following table lists the bit definitions for the Standard Event
Bit Number
0 Not used
1 Not used
2 Query Error
3 Device Error
4 Execution Error
5 Command Error
6 Not used
7 Power On
• Use the <enable_value> parameter to specify which bits will be enabled.
• The *CLS (clear status) command will not clear the enable register but
• An Instrument Preset (SYSTem:PRESet command) does not clear the
• Refer to
Example
The following command enables bit 4 (decimal value = 16) in the enable
register. If an Execution Error occurs, this condition will be reported to
the Status Byte Register (bit 5 will be set to high).
36
Register.
Decimal Value
Not used
Not used
4
Not used
16
32
Not used
128
The decimal value specified corresponds to the binary- weighted sum of
the bits you wish to enable in the register. For example, to enable bit 2
(decimal value = 4), bit 3 (decimal value = 8), and bit 7 (decimal value
= 128), the corresponding decimal value would be 140 (4 + 8 + 128).
it clears all bits in the event register.
bits in the Status Byte enable register.
Figure 6- 1
Definition
"0" is returned
"0" is returned
The instrument tried to read the output buffer but it was
empty. Or, a new command line was received before a
previous query has been read. Or, both the input and
output buffers are full
"0" is returned
An execution error occurred
A command syntax error occurred
"0" is returned
Power has been turned off and on since the last time
the event register was read or cleared
on page 37 for the status system diagram.