1
Introduction to Programming
ROUTe:CLOSe (@<ch_list>)
INPut:IMPedance:AUTO <mode> [,(@<ch_list>)]
As shown above, the <ch_list> parameter is optional for some
commands (as indicated by square brackets). If you omit the <ch_list>
parameter, the command will be applied to the internal DMM.
The following command closes channel 10 on the module in slot 3.
ROUTe:CLOSe (@3010)
The following command closes channels 10, 12, and 15 on the module in
slot 2.
ROUTe:CLOSe (@2010,2012,2015)
The following command closes channels 5 through 10 (slot 1) and channel
15 (slot 2). When you specify a range of channels, any channels that are
invalid will be ignored (no error will be generated) but the first and last
channel in the range must be valid.
ROUTe:CLOSe (@1005:1010,2015)
The Analog Bus relays (numbered s911, s912, s913, etc.) on the
multiplexer and matrix modules are ignored if they are included in a
range of channels. An error will be generated if an Analog Bus relay is
specified as the first or last channel in a range of channels. For example,
the following command closes all valid channels between channel 30 (slot
1) and channel 5 (slot 2). In addition, this command closes Analog Bus
relay 911 on the module in slot 1 (Bank 1). Note that although the
specified range of channels includes the other Analog Bus relays, they are
ignored and are not closed by this command.
ROUTe:CLOSe (@1030:2005,1911)
The following command will generate an error since the Analog Bus relays
cannot be specified as the first or last channel in a range of channels
(none of the channels will be closed).
ROUTe:CLOSe (@1005:1911) !Generates an error