8
OUTPut Subsystem

OUTPut

Syntax
OUTPut <mode>
This command sets the status (ON/OFF) of the analog output.

OUTPut?

The query command returns a string value representing the status of the
analog output.
Supported Model
U2351A, U2353A, U2355A, U2356A, U2331A
Parameters
Remarks
• The OUTPut ON and OUTPut OFF commands cannot be executed
• Resets the voltage level by using the SOURce:VOLTage 0 command.
• Cannot generate user- defined and pre- defined waveforms at the same
Query Returned Format
<mode> ::= {0|1}
64
Name
Type
<mode >
String
repeatedly.
time.
Range of Values
OFF or 0: Output predefined waveform
ON or 1: Terminate waveform output
U2300A Series USB DAQ Programming Guide
Default Value
0