Route:channel:rvoltage - Agilent Technologies U2300A Series Programming Manual

Multifunction usb data acquisition
Hide thumbs Also See for U2300A Series:
Table of Contents

Advertisement

ROUTe:CHANnel:RVOLtage

Syntax
ROUTe:CHANnel:RVOLtage <value>, (@<ch_list>)
This command is used to set the voltage value of the external reference
ENABlevoltage source of the analog outputs.
ROUTe:CHANnel:RVOLtage? (@<ch_list>)
The query command returns a numeric value that represents the external
reference voltage value in volts for analog outputs.
Supported Model
U2351A, U2353A, U2355A, U2356A, U2331A
Parameters
Item
<value>
<ch_list>
Remarks
• It will return an error code if OUTPut ON.
• The reference voltage value cannot be set to zero.
Query Returned Format
<value> [V]
U2300A Series USB DAQ Programming Guide
Type
Range of Values
Numeric
From 0 to 10, if the D/A reference voltage source is the
external reference.
Numeric
Single-ended mode:
U2351A/U2352A/U2353A/U2354A: 101-116
U2355A/U2356A/U2331A: 101-164
Differential mode:
U2351A/U2352A/U2353A/U2354A: 101-108
U2355A/U2356A/U2331A: 101-132
ROUTe Subsystem
10
Default Value
10
N/A
95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U2351aU2353aU2354aU2355aU2356aU2331a ... Show all

Table of Contents