4
CALibration Subsystem
CALibration:BEGin
Syntax
CALibration:BEGin
This command is used to calibrate your device. When the function is
executed, the device goes into a self- calibration mode. This command
initiates a voltage calibration (adjustment) sequence for the specified DAC
channel on the USB DAQ device. This sequence sets both zero and gain
adjustment constants for each of the DAC output.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Remarks
Use the
calibration is complete if the returned value is "1". If the returned value
is "0", the calibration is still in progress.
24
*OPC?
command to check if the calibration is completed. The
U2300A Series USB DAQ Programming Guide