OUTPut Subsystem

OUTPut:WAVeform:SRATe

Syntax
OUTPut:WAVeform:SRATe <value>
This command is used to set the updated rate of analog outputs.

OUTPut:WAVeform:SRATe?

The query command returns an integer value representing the updated
rate of analog outputs measured in Hz.
Supported Model
U2351A, U2353A, U2355A, U2356A, U2331A
Parameters
Remarks
• The range bound by 0 ≤ <value> ≤ 10
• Default value zero implies automatic sampling configuration for SIN,
• Maximum of 4096 points for one waveform.
Query Returned Format
<value> [Hz]
66
Name
Type
<value>
Float
SQU, SAWT, TRI, and NOIS waveform generation.
Range of Values
(48/16777215) MHz to 1 MHz
. Must be ≥ 48x10
6
U2300A Series USB DAQ Programming Guide
Default Value
0
6
/ 0xffffff.