Performance Tests
Table C-2
Deflection Accuracy Limits
VOLTS/DIV
STANDARD
ACCURACY
switch
amplitude
limits
setting
signal
(divisions)
2 mV
10 m V
4.85 to 5.15
5 mV
2 0 mV
3.88 to 4.12
10 mV
50 mV
4.85 to 5.15
20 mV
0.1 V
4.85to5.15
50 mV
0.2 V
3.88 to 4.12
0.1 V
0.5 V
4.85to5.15
0.2 V
1 V
4.85 to 5.15
0.5 V
2V
3.88 to 4.12
1V
5V
4.85 to 5.15
2V
10V
4.85 to 5.15
5V
20 V
3.88 to 4.12
Step 3.
Check Non-Store (Analog) Bandwidth
a. SET:
VOLTS/DIV (CH 1 and CH 2)
5 mV
Vertical MODE CH
1
SEC/DIV
10
µs
b. Connect the leveled sine-wave generator output via a
50 Q
BNC coaxial cable and a
50 Q
BNC termination to
the
CH 1 OR X
input connector.
c. Set the generator to produce a
50
kHz, six-division
display.
d. Increase the signal frequency until a
4.2
division
display is obtained.
e. Check that the frequency is greater than
60
MHz.
f.
Repeat parts b through e for all
VOLTS/DIV
settings
from
10
mV through
1
V.