Performance Tests
Equipment Required (see Table B1):
Calibration Generator (Item 1)
Leveled Sine-Wave Generator (Item 2)
50
n
ENC Coaxial Cable (Item 7)
Dual Input Coupler (Item 8)
50
n
ENC Termination (Item 11)
lOX ENC Attenuator (Item 13)
ENC Male-to-Tip Plug (Item 14)
Initial Control Settings
POSITION ..................................... Midrange (CH1 and CH2)
VOLTS/DIV ................................... 2 mV (CHl and CH2)
VARiable ....................................... Off (CHl and CH2)
AC-GND-DC .................................. DC (CHl and CH2)
POSITION ..................................... Midrange
MODE ALT MAGN ....................... Off
SEC/DIV ......................................... 0.5 ms
V ARiable ....................................... Off
X1, XlO, X50 Magnifier ................ Xl
HOLDOFF ...................................... MIN
LEVEL ............................................ Midrange
SLOPE ............................................ Positive Going
SOURCE ........................................ VER Ti cal MODE
2212 User Manual
C-9