X-Y Display Checks; Check Probe Adjust Operation - Tektronix 2212 User Manual

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Procedure Steps
Step
1.
Check Probe Adjust Operation
a. SET:
VOL TS/DIV (CH
1)
SEC/DIV
0.5
ms
Trigger SOURCECH
1
10mV
Performance Tests
b. Connect the
lOX
probe to the CH 1 OR X input con-
nector and clip the probe tip to the PROBE ADJUST con-
nector on the front panel.
c. If necessary, adjust the probe compensation for a flat-
topped squarewave display (See also Figure
2-1
on
page
2-4).
d. CHECK -That the display amplitude is between
4.75
to
5.25
divisions.
X-Y Display Checks
The
X-Y
Display Check procedures check those characteristics
that relate to the
X-Y
Display system and that are listed under
Warranted Characteristics in Appendix B: Specifications. You
should set up the test equipment as shown at the start of the pro-
cedure list. Changes to the test set-up will be indicated in the pro-
cedures, if necessary.
2212 User Manual
C- 27