Vertical Store (Digital) Display Mode; Figure 5-2: Vertical Controls And Connectors - Tektronix 2212 User Manual

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Vertical
5-8
• Vertical Store (Digital) display mode.
With the 2212 in the Store display mode, you can select:
CH 1, ADD, ALT, CHOP, X-Y or CH 2.
If
the ADD, ALT or CHOP mode is selected in Store mode, the
signals applied to the vertical input connectors are acquired
simultaneously. However, if the vertical mode is in ALT and
the trigger source is in VERT MODE, CH 1 and CH 2 will be
digitized alternately.
In the X-Y Store mode, the acquisition and the sampling rate
are set by the SEC/DIV switch and/ or the EXT CLOCK.
15
1IPOSITION
v(Q)
CH 1
VOLTS/DIV
CHlOR[!)
25pF
lMO
0
~4~Vpk
0
16
VERTICAL
I
1IPOSITION
MODE
V
©
-----i
11
CH2oR00
25pF
lMO
8
~4~Vpk
0
Figure 5-2: Vertical Controls and Connectors