Table C-1: Test Equipment Required - Tektronix 2212 User Manual

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Performance Tests
Table C-1: Test Equipment Required
Item and
Minimum
Example of
Description
Specification
Use
Test Equipment
1. Calibration
Standard-amplitude sig- Signal source
Tektronix
Generator
nal levels: 5 mV to 50 V. for gain and
PG 506A
Accuracy: ±0.3 %.
transient
Calibration
- High-amplitude signal
response
Generator. •
levels: 1 V to 60 V.
checks and
Repetition rate: 1 kHz.
adjustments.
- Fast rise signal level:
1 V. Repetition rate:
1 MHz. Risetime: 1 ns
or less. Flatness: ±0.5%.
2. Leveled
Frequency: 250 kHz to
Vertical,
Tektronix
Sine-Wave
above 60 MHz.
horizontal, and
SG503 Leveled
amplitude
Output:: variable from
triggering
Sine-Wave
Generator
10 mV to 5V p-p. Out-
checks and
Generator. •
put impedance: 50
n.
adjustments.
Reference frequency:
Display adjust-
50 kHz. Amplitude
ments and
accuracy: constant with- Z-Axis check.
in 3% of reference
frequency as output
frequency changes.
3. Time-Mark
Marker outputs: 10 ns to Horizontal
Tektronix
Generator
0.5 s. Marker accuracy:
checks and
TG501 Time-
±0.1 %. Trigger output:
adjustments.
Mark Genera-
1 ms to 0.1 µs, time-
Display
tor
a
coincident with
adjustment
markers.
2212 User Manual
C-3