Non-Storage Examples
Time Difference Measurement
Between Two Time-Related Pulses
The calibrated sweep speed and dual-trace features of the instru-
ment allow measurement of the time difference between two sepa-
rate events. To obtain a reliable measurement result, use either
probes or cables with equal time delays. You can use the following
procedure to measure time difference (see Figure
3-7):
Step
1.
Switch the 2212 on and obtain a baseline display in
Non-Store mode.
LiT
~4.50
100 · ·
....
90
10
0%" ••
..
..
2V
I
I
INACTIVE
CURSOR
111S.
T~
. i"
.... .. . .
:
"i"
....
....
1V
AC
:
'
... ....
.... . ...
5111S
,,
V~RT
.
... . ..
'~
REFERENCE
SIGNAL
,llf
.
..
'
:-:-:-:
'
ACTIVE
CURSOR
COMPARISON
SIGNAL
Figure 3-7: Time Difference Measurement between two Time-
related Pulses
Step 2.
Toggle the Trigger SOURCE switch to CH
1.
Step
3.
Set both AC-GND-DC switches to the same position,
depending on the type of input coupling desired (AC or
DC).