Setup
SETUP Controls
5 - 42
The AUTO SETUP feature enables you to automatically setup the
horizontal, the vertical, and trigger controls in order to produce a
display that fits vertically and horizontally on the screen, with at
least two periods of the probed signal displayed. A front panel
setup can be saved in memory and recalled from that memory. The
saved front panel setup is retained in memory even after the in-
strument is switched off. See APPENDIX A for the default control
settings of each front panel control at power up.
See Figure 5-7 for locations of items 46 through 49.
49
SETUP
SAVE
RECALL
QMENU9
Figure 5-7: SETUP controls
@
AUTO Setup Switch
The AUTO Setup switch selects the AUTO SETUP function.
When you press the AUTO button, the vertical, horizontal,
triggering and display controls to display an unknown
waveform are automatically set. A stable, automatically
triggered display of the probed waveform appears on-screen
with an optimized front panel setup.
Auto Setup does not affect the STORAGE and CURSOR
modes and controls.
In Detail

Hide quick links: