Tektronix 2212 User Manual page 206

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Performance Tests
Table C-1 (cont.)
Test Equipment Required
Item and
Minimum
Example of
Description
Specification
Use
Test Equipment
4. Low-
Range 10 Hz to
Low-
Tektronix
Frequency
500 kHz.Output ampli-
Frequency
SG502
Sine-Wave
tude: 300 mV. Output
trigger
Oscillator.•
Generator
impedance: 600
n.
Ref-
checks
erence frequency:
constant within 0.3 dB
of reference frequency
as output frequency
changes.
5. Pulse
Repetition rate:
Signal source
Tektronix
Generator
1 kHz. Output
for Storage
PG501
amplitude: 5 V.
and external
Pulse Genera-
clock checks.
tor.
a
6. TV Signal
Provide Composite
Check TV
Tektronix
Generator
TV Video and Line
Trigger circuit
TSG-100
Sync Signals
Test Signal
Generator.
7. Coaxial
Impedance: 50
n.
Signal inter-
Tektronix Part
Cable (2x)
Length: 42 in.
connection.
Number
Connectors: BNC.
012-0057-01.
8. Dual Input
Connectors: BNC.
Signal inter-
Tektronix Part
Coupler
Female-to-dual-
connection.
Number
BNC male.
067-0525-02.
9. Precision
Impedance: 50
n.
Vertical Band-
Tektronix Part
Coaxial
Length: 42 in.
width
Number
Cable
Connectors: BNC.
012-0482-00.
10. T-Connector
Connector: BNC.
Signal inter-
Tektronix Part
connection.
Number
103-0030-00.
11. Termination
Impedance: 50
n.
Signal
Tektronix Part
Connectors: BNC.
Termination.
Number
011-0049-01.
C-4
Appendix C: Performance Verification