Performance Tests
k. Adjust the Trigger
LEVEL
control so that the sweep
starts two divisions above the center horizontal graticule
line and check that the trigger readout is 0.20 V ±0.03 V.
For three divisions it is 0.30 V ± 0.03 V.
1. Adjust the Trigger
LEVEL
control so that the sweep
starts one division below the center horizontal graticule
line and check that the trigger readout is - 0.10 V ±0.03
V. For two divisions it is - 0.20 V± 0.03 V and for three
divisions - 0.30 V ± 0.03 V.
m. Disconnect the test equipment from the instrument.
Step 8.
LINE Trigger Function Check
a. SET:
CH 2 VOLTS/DIV .................... 0.1 V (without l0X probe
attached)
CH 2 AC-GND-DC .................. DC
SEC/DIV ................................... 5 ms
Trigger MODE ......................... P-P AUTO
Trigger SOURCE ...................... LINE
Trigger COUPLING ................. DC
Xl, XlO, X50 Magnify ............. Xl
ALT MAGN switch ................. Off
b. Connect a l0X probe to the
CH
2
ORY
input connec-
tor.
c. Attach a length of wire of two inches long to the
probe tip. Hold the wire near the middle portion of the
instrument power cord.
d. CHECK - That the display can be triggered on posi-
tive-going and negative-going slopes.
e. Disconnect the test set up from the instrument.